CONG Prima-10 (in flow-through configuration)

CONG Prima-10

The CONG Prima-10 dew point analyzer is an automatic on-line hygrometer and a supporting secondary unit and is designed to measure the water dew point and hydrocarbon condensation temperature in natural gas.
This analyzer is designed to measure the dew point of water even in the presence of substances that condense earlier (glycols, methanol, higher hydrocarbons).
The analyzer consists of an explosion-proof dew point transducer (1Ex d IIA T5 Gb) and a secondary unit. There are two designs for the secondary unit: a central control unit (CCU) for installation in a 19-inch instrument rack and an interface unit (IU) for attachment to a vertical panel. The IU/CCU provides registration and storage of measurement data, conversion of measured dew point values to moisture content (g/m³), and converts natural gas water dew point values from the measured pressure to the contract pressure. By connecting the IU/CCU to an external computer it is possible to monitor on-line dew point measurements in order to evaluate measurement reliability.
The transducer can be mounted directly onto a gas pipeline (submersible configuration) or connected to the pipeline through a gas preparation system with a filtration option for the removal of mechanical and aerosol contaminants.

Fields of application

  • gas
  • oil
  • chemical
  • power
  • metallurgical industry

Signature
features

Automatic diagnostics of the cooling mirror
The analyzer’s operational algorithm includes automatic diagnostics of the cooling mirror surface and automatic cleaning of the mirror before each measurement cycle.
Increased difference between body temperature and the minimum mirror surface temperature.
The next-generation sensors integrated into the instrument’s innovative design make it possible to achieve temperature depression of 70 °C between the analyzer’s body temperature and the minimum temperature of the mirror’s surface at an operating pressure of 10 MPa. This is the best result among all known automatic chilled-mirror on-line hygrometers.

Characteristics

Type
Laser-base, condensation mirror, automatic on-line, stationary
Water dew point measurement range
from -50 to +30°С
Hydrocarbon condensation temperature measurement range
from -30 to +30°С
Measurement chamber gas flow rate
1...2 l/min
Error rate limits for dew point measurements:
±0,25°С
±1°С
Maximum operating pressure of target gas:
up to 16 MPa
up to 25 MPa
IEC 60529 degree of protection
IP67
Explosion protection
1 Ex d IIA T5Gb
II 2 G Ex d IIA T5 Gb
Connection to supply line of target gas
Swagelok connector for a pipe with outside diameter of 6mm
Power supply voltage
~ 220 V
Output signals:
Digital: RS-232/RS-485 protocol Modbus/RTU
Analog: 2 outputs (4...20) mA
Measurement cycle time:
from 5 to 10 min
Dimensions, max:
see attachment Г, Д, М4, М6 КРАУ2.844.005-03(-04) РЭ
(depends on complectation)
Weight kg,max:
6,5 kg
ПТР с СПГ КРАУ2.848.012(-01,-02): 38,5
ИБ КРАУ3.035.001-003
ЦУБм КРАУ: 6,5
Шкаф специализированный всепогодный: 400
show all Characteristics

Configuration*

show package contents
Main set
KRAU2.848.004 (KRAU2.848.004-01, KRAU2.848.004-02) dew point transducer (the design is specified at the time of order)
1
KRAU3.622.002-01 interface unit
1
gas purification system 
KRAU2.848.012 (-01, -02)
1
KRAU2.848.013
1
KRAU4.078.182 (KRAU4.078.183, KRAU4.078.184, KRAU4.078.185) gas purification system connection kit (design specified at the time of order)
1
Software (on CD):
"Trassa-2"
1
«Hygrovision»
1
«ОРС Server CONG-Prima» software (CD + Guardant dongle + User manual)
included by order
Portable workstation with special software
included by order
Operational documents
1
KRAU2.844.005 МП calibration methods
1
* may be altered. Clarified at the time of order.
В состав комплекта входят: шкаф специализированный всепогодный ТШВ-1795, пробоотборное устройство КРАУ6.457.013, фильтр КРАУ6.457.176, пробоотборная линия, диэлектрическая вставка SS-12-MDE-6, элементы обогрева пробоотборной линии и СПГ
show package contents

Model 003 gas preparation system

Automatic diagnostics of the cooling mirror
The analyzer’s operational algorithm includes automatic diagnostics of the cooling mirror surface and automatic cleaning of the mirror before each measurement cycle.
Increased difference between body temperature and the minimum mirror surface temperature.
The next-generation sensors integrated into the instrument’s innovative design make it possible to achieve temperature depression of 70 °C between the analyzer’s body temperature and the minimum temperature of the mirror’s surface at an operating pressure of 10 MPa. This is the best result among all known automatic chilled-mirror on-line hygrometers.